7 Analog Modules Time Interval Analyzer
226 System Reference, January 2001
Figure 101 TIA Measurement Functions
The TIA can measure intervals with the specified numbers
of samples and store them into the local memory, then
return the following data calculated from the raw sampled
data as the measurement results:
•Maximum value
• Minimum value
• Jitter RMS (standard deviation)
• Mean (averaged value)
Edges to Start and Stop Measurement
For edge types determined by the
measureMode
setting,
the TIA starts counting edges for start and/or stop chan-
nels to perform the time measurement after the trigger
signal is detected.
trigDelayCntStart
and
trigDe-
layCntStop
are used to specify at which edges to start
and stop the measurement. They have different meanings
depending on which measurement is performed:
CH 1
Period (PER)
= 1/Frequency (FREQ)
Pulse Width (PW+)
Pulse Width (PW-)
CH 1
Start Channel
CH 2
Stop Channel
PD+-
PD-+ PD--
Propagation Delay
PD++
Single Channel
Dual Channel
tia_ov1