Synchronization 7 Analog Modules
System Reference, January 2001
249
Adjusting Synchronization Timing
Considering Trigger-to-Signal Delay Only
If the trigger line delay and the signal line delay are quite
smaller than the trigger-to-signal delay, you can ignore
trigger line length and signal line length. Suppose you are
using a digitizer to measure an analog signal from a DUT.
You have to program the edge location of the trigger signal
before the timing when the analog signal is output from
the DUT by the trigger-to-signal delay. The following is the
timing chart.
Figure 116 Adjusting Synchronization Timing (1)
Measurement start
@DUT pin
tgtdl
Analog signal
@DUT pin
Trigger-to-signal
delay
Trigger edge setting
Move backward by
trigger-to-signal delay